Test Code MENES Environment and Sterility Test (Endoscope)
Laboratory Code
LAB3053139
Specimen Requirements
Specimen Type: Automatic Endoscope Reprocessors (AER) water, Endoscope flush fluid
Container/Tube: Sterile screw-capped container
Volume: AER collection: at least 200.0 mL
Endoscope flush fluid: 50.0 – 200.0 mL
Minimum volume: 50.0 mL
Collection Instructions: As described in The National Infection Prevention and Control Guidelines for Endoscopy Centres 2023.
Specimen Stability
Specimen Type | Water |
---|---|
Ambient | Send specimen to the laboratory immediately, to be received in testing laboratory within 1 hour of collection |
Refrigerate | To be received in testing laboratory within 6 hours of collection (preferred for non-NUH requests) |
Frozen | Unacceptable |
Methodology
Culture
Days and Times Performed
Monday through Saturday, TAT 8 days
- Preliminary result issued in 2 days and final report on Day 7.
- Sample to reach Microbiology Laboratory by 4 pm on test day (for non-NUH requests).
Reference Interval
No bacterial growth after 7 days' incubation.
Sterility satisfactory.
Testing Laboratory
National University Hospital
Clinical Microbiology, 67724343
Useful For
Monitoring the effectiveness of cleaning and disinfection process of endoscopes.
Service Code
23083
LOINC Code Information
Not applicable
Update
Last updated 21/3/2025
Specimen Conditions
Insufficient specimen will be rejected.