Test Code LAGGN Granulocyte Antibody Screen, Serum
Additional Codes
| Epic Ordering Code | LAB305380 |
| NBS code | LBC13485 |
| Service code | 37224 |
Reporting Name
Granulocyte Ab Screen, SSpecimen Type
Serum RedSpecimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Red top (serum gel/SST are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 1.5 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Additional Information: Only a specimen collected before a transfusion reaction is acceptable.
NUH Outreach Laboratories:
Specimen Requirements
Submit the following:
Container: Red Top
Only a specimen collected before a transfusion reaction is acceptable.
Processing Requirements
Centrifuge and aliquot serum into a Mayo plastic tube.
Specimen type: Serum
Volume: 1.5 mL
Storage/transport requirements: Refrigerated
Specimen Minimum Volume
0.3 mL
Specimen Stability Information
| Specimen Type | Temperature | Time |
|---|---|---|
| Serum Red | Refrigerated (preferred) | 30 days |
| Frozen | 365 days | |
| Ambient | 7 days |
Reject Due To
| Gross hemolysis | Reject |
| Gross lipemia | OK |
| Gross icterus | OK |
Method Name
Flow Cytometry/Agglutination
Day(s) Performed
Tuesday, Thursday
Report Available
7 to 15 daysPlease note that this is a referral test requiring transport to the external testing facility and an additional 3-5 days is required.
Reference Values
Not applicable
Performing Laboratory
Mayo Clinic Laboratories in Rochester
Useful For
Work-up of individuals with autoimmune neutropenia
Work-up of individuals having febrile nonhemolytic transfusion reactions
Work-up for alloimmune neonatal neutropenia
This test is not useful for the diagnosis of neutropenia due to marrow suppression by drugs or tumors.
CPT Code Information
86021 x2
LOINC Code Information
| Test ID | Test Order Name | Order LOINC Value |
|---|---|---|
| LAGGN | Granulocyte Ab Screen, S | 105285-1 |
| Result ID | Test Result Name | Result LOINC Value |
|---|---|---|
| LAGG2 | GIFT/GAT Interpretation | 105288-5 |
| LAGG3 | GIFT Result | 105286-9 |
| LAGG4 | GAT Result | 105287-7 |
Secondary ID
621630Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.Last updated 28/07/2026