Sign in →

Test Code IHS Coagulation Inhibitor Screen

Important Note

Please consult haematologist before ordering this test.

Please do not order this test for patients with normal PT and APTT results, unless clinically indicated.

Specimen Requirements

Submit only one of the following specimens


Specimen Type: Whole Blood
Container/Tube: 3 Light Blue Top (3.2% Sodium Citrate)
Volume: 3 × 2.7mL Whole Blood (Tubes must be full)
Collection Instructions: Send whole blood specimens to laboratory immediately.


Specimen Type: Platelet Poor Plasma
Container/Tube: 3 Light Blue Top (3.2% Sodium Citrate)
Submission Container/Tube: Plastic Vial

Minimum Volume: 3.0mL Plasma
Collection Instructions:

1. Light blue top tube must be properly filled with whole blood.
2. Whole blood specimen should be double centrifuged and plasma separated from cells within 1 hour of collection.
3. Plasma should be separated into 3 aliquots with minimum volume of 1mL each.
4. Freeze the aliquots in well-closed plastic tubes immediately.

Reference Interval

Absent

Days and Times Performed

Monday through Friday, Office hours, TAT 5 day

Methodology

APTT-Based Coagulation Test

Service Code

12412

Update

Last updated 25/02/2020

Unacceptable Condition

Haemolysis Gross reject
Icterus NA
Lipaemic NA
Others Clotted or insufficient reject

 

Specimen Stability

**Preferred temperatures for storage prior to and during shipping to NUH**

 

Specimen Type Plasma
Ambient 4 hours
Refrigerate NA
Frozen 14 days

 

Testing Laboratory

National University Hospital
Haematology, 67725320

Useful For

Aid in the detection of circulating lupus anticoagulant.

Aid in the detection of coagulation inhibitors, e.g. Acquired Factor VIII inhibitor.

Alias

Inhibitor Screen

Non-specific Inhibitor Screen