Sign in →

Test Code 25OHD Vitamin D 25-Hydroxy

Laboratory Code

LAB535

Specimen Requirements

Specimen Type: Serum
Container/Tube: Gold Top (SSTII)
Volume: 5.0 mL
Minimum Volume: 0.5 mL serum
Collection Instructions: Allow specimen to clot for 30 mins and centrifuge within 2 hrs.

Specimen Stability

Specimen Type Serum
Ambient 72 hours
Refrigerate 12 days
Frozen 1 year

Methodology

Chemiluminescent Immunoassay

Days and Times Performed

24 hrs, TAT 1 day

Specimen Condition

Gross Haemolysis Alert/Reject
Icterus OK
Lipaemia Alert/Reject
Others Not applicable

Reference Interval

Deficient  ≤ 20 ug/L
Insufficient 21 – 29 ug/L
Sufficient 30 – 100 ug/L
Possible toxicity  > 100 ug/L

Testing Laboratory

National University Hospital

Clinical Chemistry, 67724346

Useful For

Diagnosis of vitamin D deficiency and hypervitaminosis D.

Monitoring vitamin D replacement therapy

Service Code

31036

LOINC Code Information

1989-3

Alias

25-OH Vitamin D2 and D3, Serum

Update

Last updated 13/03/2025